In:
Clinical Chemistry, Oxford University Press (OUP), Vol. 32, No. 3 ( 1986-03-01), p. 532-535
Abstract:
A chemiluminescent labeled-antibody immunoassay has been developed for measurement of total thyroxin (T4) in serum. Monoclonal antibodies to T4 labeled with a chemiluminescent acridinium ester are used. Serum samples are incubated with the labeled antibodies and a thyroxin-rabbit immunoglobulin conjugate, then reacted with magnetizable particles containing sheep anti-rabbit immunoglobulin. The total reaction time is 40 min. The chemiluminescence intensity of the solid-phase immune complexes is inversely proportional to the concentration of T4 in the sample. The sensitivity of the assay is 1 nmol/L, and the working range of 20-190 nmol/L is characterized by CVs less than or equal to 10%.
Type of Medium:
Online Resource
ISSN:
0009-9147
,
1530-8561
DOI:
10.1093/clinchem/32.3.532
Language:
English
Publisher:
Oxford University Press (OUP)
Publication Date:
1986