In:
Viruses, MDPI AG, Vol. 14, No. 5 ( 2022-05-17), p. 1065-
Abstract:
To evaluate the diagnostic performance of the Liaison® Murex anti-HEV IgM and IgG assays running on the Liaison® instrument and compare the results with those obtained with Wantai HEV assays. We tested samples collected in immunocompetent and immunocompromised patients during the acute (HEV RNA positive, anti-HEV IgM positive) and the post-viremic phase (HEV RNA negative, anti-HEV IgM positive) of infections. The specificity was assessed by testing HEV RNA negative/anti-HEV IgG-IgM negative samples. The clinical sensitivity of the Liaison® IgM assay was 100% for acute-phase samples (56/56) and 57.4% (27/47) for post-viremic samples from immunocompetent patients. It was 93.8% (30/32) for acute-phase (viremic) samples and 71%% (22/31) for post-viremic samples from immunocompromised patients. The clinical sensitivity of the Liaison® IgG assay was 100% for viremic samples (56/56) and 94.6% (43/47) for post-viremic samples from immunocompetent patients. It was 84.3% (27/32) for viremic samples and 93.5% (29/31) for post-viremic samples from immunocompromised patients. Specificity was very high ( 〉 99%) in both populations. We checked the limit of detection stated for the Liaison® IgG assay (0.3 U/mL). The clinical performance of the Liaison® ANTI-HEV assays was good. These rapid, automated assays for detecting anti-HEV antibodies will greatly enhance the arsenal for diagnosing HEV infections.
Type of Medium:
Online Resource
ISSN:
1999-4915
Language:
English
Publisher:
MDPI AG
Publication Date:
2022
detail.hit.zdb_id:
2516098-9