Journal of Pharmacy Technology, SAGE Publications, Vol. 16, No. 1 ( 2000-01), p. 12-17
Abstract:
To determine whether storage at low temperatures affects the drug delivery efficacy of a chlorofluorocarbon (CFC)-free aerosol and a dry-powder inhaler (DPI). Design: Aerosol output from a CFC-free hydrofluoroalkane (HFA-134a) metered-dose inhaler (albuterol) and a multidose DPI (terbutaline) were examined at different ambient temperatures and canister loads. Results: After the HFA-134a formulation was stored at low temperatures for 24 hours, a small decrease of aerosol output (2.1–2.7% at 0 °C, 5.0–8.0% at −10 °C; p 〈 0.05 from 22 °C) was observed. In addition, the aerosol output from devices with one-half or one-fifth the initial content was marginally decreased compared with output from inhalers with full content. Image analysis techniques applied to fully formed aerosol plume emitted from the HFA-134a formulation showed that the product could still be aerosolized at 10 °C, but there was a reduction of the distance reached by the aerosol. In contrast, condensed droplets markedly altered the aerosols formed from CFC inhalers at −10 °C. Aerosol output from the DPI apparently was not altered following 24-hour storage at −10 °C. Conclusions: Aerosols emitted by the HFA-134a and DPIs studied were less affected by exposure to cold temperatures compared with those from marketed CFC devices.