Abstract:
Air entrapment (AE) has been reported as a potential cause of early inappropriate shocks (ISs) following subcutaneous implantable cardioverter defibrillator (S-ICD) implantation, but a cause–effect relationship is not always evident. This systematic review aims to analyse this phenomenon concerning implantation techniques, electrogram (EGM) features, radiologic findings, and patient management. Methods and results A systematic search was conducted using PubMed, Embase, and Google Scholar databases following the PRISMA guidelines to obtain all available literature data since 2010 on S-ICD malfunctions possibly due to AE. The final analysis included 54 patients with AE as a potential cause of S-ICD malfunction. Overall, the aggregate incidence of this condition was 1.2%. Of ICD malfunctions possibly due to AE, 93% were ISs, and 95% were recorded within the first week following implantation. Radiologic diagnosis of AE was confirmed in 28% of the entire study cohort and in 68% of patients in whom this diagnostic examination was reported. At the time of device malfunction, EGMs showed artefacts, baseline drift, and QRS voltage reduction in 95, 76, and 67% of episodes, respectively. Management included ICD reprogramming or testing, no action (observation), and invasive implant revision in 57, 33, and 10% of patients, respectively. No recurrences occurred during follow-up, irrespective of management performed. Conclusions Device malfunction possibly due to AE may occur in ∼1% of S-ICD recipients. Diagnosis is strongly suggested by early occurrence, characteristic EGM features, and radiologic findings. Non-invasive management, principally device reprogramming, appears to be effective in most patients.
Type of Medium:
Online Resource
ISSN:
1099-5129
,
1532-2092
DOI:
10.1093/europace/euac046
Language:
English
Publisher:
Oxford University Press (OUP)
Publication Date:
2022
detail.hit.zdb_id:
2002579-8
Bookmarklink