Abstract:
Detailed analysis of the thermoluminescence (TL) dosimetry method for ß radiation has been performed based on determination of the correction factor for energy dependence using the ratio J12 of TL responses of two detectors with different thickness (H1 & lt; H2). It has been found that the optimum PVC filter thickness on the thick detector enhancing the resolution of the method is equal to 0.005-0.010 cm. It has been shown that the angular distribution of ß radiation causes a negligible error in the determination of the correction factor.