In:
Clinical Chemistry, Oxford University Press (OUP), Vol. 39, No. 7 ( 1993-07-01), p. 1435-1439
Abstract:
The influence of assay design and quantification system on assay performance was investigated by developing, optimizing, and comparing a time-resolved immunofluorometric assay (IFMA), an immunoenzymometric assay (IEMA), an immunoradiometric assay (IRMA), and a competitive radioimmunoassay (RIA), all performed with the same monoclonal antibodies (MCA) directed against human follicle-stimulating hormone (hFSH). The lowest detection limit (2 ng/L for hFSH-I-3, corresponding to 2.5 mIU of 1st International Reference Preparation of hFSH 78/549 per liter), the widest measuring range (2-160,000 ng/L), and the greatest signal-to-noise ratio (13,000:1 at 160,000 ng/L) were obtained in the IFMA. For analysis of serum samples from 101 male (ages 2-91 years) and 99 female (ages 2-90 years) individuals at a single dilution, 100% of samples were within the measuring range of the IFMA, whereas only 87%, 55%, 32%, and 8% of the sera were for the IRMA, the IEMA evaluated with double-wavelength measurement, the conventional IEMA, and the competitive MCA-based RIA, respectively. These studies demonstrate clear advantages of the IFMA in sensitivity and assay range, which allows reliable and cost- and time-effective determination of hFSH in individuals from infancy to senescence.
Type of Medium:
Online Resource
ISSN:
0009-9147
,
1530-8561
DOI:
10.1093/clinchem/39.7.1435
Language:
English
Publisher:
Oxford University Press (OUP)
Publication Date:
1993