In:
Journal of Endovascular Therapy, SAGE Publications, Vol. 14, No. 2 ( 2007-04), p. 260-263
Abstract:
To report a novel complication of a hybrid “frozen elephant trunk” endograft used in the treatment of a multisegmentally diseased aorta. Case Report: A 53-year-old man with chronic type A aortic dissection and previous replacement of the supracoronary ascending aorta underwent a “frozen elephant trunk” procedure using a hybrid endograft. The stent-graft was placed deep in the descending aorta to cover the multiple distal re-entries. Due to ineffective covering of the re-entries, a patent false lumen led to rapid repressurization of the false lumen immediately distal to the circumferential elephant trunk anastomosis. Compression of the non-stented portion of the hybrid endograft caused a functional aortic stenosis, with severe hemodynamic consequences. Endovascular treatment of the compression by retrograde transfemoral placement of a stent-graft in the non-stented portion of the hybrid endograft achieved free flow in the distal aorta. Conclusion: This case documents a new complication of the frozen elephant trunk procedure; a pseudocoarctation from a repressurized proximal false lumen was successfully managed with a stent-graft to support the non-stented segment of the hybrid endograft.