Abstract:
We reviewed our institutional experience with outflow graft stenosis (OGS) in 3 contemporary left ventricular assist devices (LVAD). METHODS Data from 347 consecutive adult recipients of LVAD [Medtronic HVAD (n = 184, 53.0%), Abbott HeartMate II (n = 62, 17.9%) and Abbott HeartMate 3 (n = 101, 29.1%)] implanted between March 2006 and October 2019 were analysed retrospectively. Primary study end points were the incidence of OGS necessitating treatment and survival on LVAD support. RESULTS During the study period, 17 patients (4.9%) developed OGS requiring treatment with a probability of 0.6% at 1 year, 1.9% at 2 years, 3.8% at 3 years, 4.7% at 4 years and 5.9% at 5 years of LVAD support. Notably, in 13.8% of patients, a compression-related narrowing of the outflow graft with a probability of 1.5% at 6 months, 1.8% 1 year, 6.0% at 2 years, 12.3% at 3 years, 15.4% at 4 years and 16.6% at 5 years of LVAD support with no difference between devices (P = 0.26) was observed. There was a trend towards increased risk of mortality with OGS (hazard ratio 2.21, 95% confidence interval 0.87–5.51; P = 0.09). OGS preferentially occurred in segments of the outflow graft covered by a protective coating. CONCLUSIONS OGS is a rare but potentially lethal complication during LVAD support. Modifications of pump design and implant techniques may be needed because OGS preferentially occurs within covered portions of the outflow graft. Systematic screening may be warranted.
Type of Medium:
Online Resource
ISSN:
1010-7940
,
1873-734X
DOI:
10.1093/ejcts/ezab382
Language:
English
Publisher:
Oxford University Press (OUP)
Publication Date:
2022
detail.hit.zdb_id:
1500330-9
Bookmarklink