In:
Journal of Clinical Microbiology, American Society for Microbiology, Vol. 48, No. 11 ( 2010-11), p. 3990-3996
Abstract:
Urinary tract infection (UTI) is a widespread disease, and thus, the most common samples tested in diagnostic microbiology laboratories are urine samples. The “gold standard” for diagnosis is still bacterial culture, but a large proportion of samples are negative. Unnecessary culture can be reduced by an effective screening test. We evaluated the performance of a new urine cytometer, the Sysmex UF-1000i (Dasit), on 703 urine samples submitted to our laboratory for culture. We compared bacteria and leukocyte (WBC) counts performed with the Sysmex UF-1000i to CFU-per-milliliter quantification on CPS agar to assess the best cutoff values. Different cutoff values of bacteria/ml and WBC/ml were compared to give the best discrimination. On the basis of the results obtained in this study, we suggest that when the Sysmex UF-1000i analyzer is used as a screening test for UTI the cutoff values should be 65 bacteria/ml and 100 WBC/ml. Diagnostic performance in terms of sensitivity (98.2%), specificity (62.1%), negative predictive value (98.7%), positive predictive value (53.7%), and diagnostic accuracy (73.3%) were satisfactory. Screening with the Sysmex UF-1000i is acceptable for routine use. In our laboratory, we have reduced the number of bacterial cultures by 43%, speeded up their reporting, and decreased the inappropriate use of antibiotics.
Type of Medium:
Online Resource
ISSN:
0095-1137
,
1098-660X
DOI:
10.1128/JCM.00975-10
Language:
English
Publisher:
American Society for Microbiology
Publication Date:
2010
detail.hit.zdb_id:
1498353-9
SSG:
12
Bookmarklink