Format:
1 Online-Ressource (33 Seiten)
ISSN:
1867-8610
Content:
We present the application of Time-of-Flight Mass Spectrometry (TOF MS) for the analysis of halocarbons in the atmosphere, after cryogenic sample preconcentration and gas chromatographic separation. For the described field of application, the Quadrupole Mass Spectrometer (QP MS) is the state-of-the-art detector. This work aims at comparing two commercially available instruments, a QP MS and a TOF MS with respect to mass resolution, mass accuracy, sensitivity, measurement precision and detector linearity. ...
Note:
Systemvoraussetzungen: PDF-Reader
In:
Atmospheric measurement techniques discussions, Katlenburg-Lindau : Copernicus, 2008-, Band 7 (2014), Seite 12323-12355, 1867-8610
In:
volume:7
In:
year:2014
In:
pages:12323-12355
Language:
English
DOI:
10.5194/amtd-7-12323-2014
URN:
urn:nbn:de:hebis:30:3-372499
Bookmarklink