A Multilaboratory Comparison of Calibration Accuracy and the Performance of External References in Analytical Ultracentrifugation

Fig 9

Histogram and box-and-whisker plot of s-values of the BSA monomer after different corrections: Raw experimental s-values (black, with grey histogram), scan time corrected st-values (blue), rotor temperature corrected s20T-values (green), or radial magnification corrected sr-values (cyan), and fully corrected s20T,t,r,v-values (red with red histogram).

The box-and-whisker plots indicate the central 50% of the data as solid line and draw the smaller and larger 25% percentiles as individual circles. The median for each group is displayed as a vertical line.

Fig 9