Evaluation of the BD Max MRSA XT assay for use with different swab types

J Clin Microbiol. 2014 Dec;52(12):4343-6. doi: 10.1128/JCM.02306-14. Epub 2014 Sep 17.

Abstract

We evaluated the performance of the BD Max MRSA XT assay for use with different swab types. The 90% detection rates (95% confidence intervals) were 387 (97 to 1,551), 877 (238 to 3,230), 986 (183 to 5,287), 1,292 (328 to 5,078), 2,400 (426 to 13,518), and 5,848 (622 to 55,021) CFU/swab for Liquid Stuart, Liquid Amies, dry, Amies Gel without charcoal, ESwab collection, and Amies gel with charcoal swabs (Becton Dickinson), respectively. Amies Gel without charcoal, ESwab collection, and Amies gel with charcoal swabs had a tendency to be less sensitive, but none of the differences was statistically significant.

Publication types

  • Evaluation Study

MeSH terms

  • Carrier State / diagnosis*
  • Carrier State / microbiology
  • Humans
  • Methicillin-Resistant Staphylococcus aureus / isolation & purification*
  • Molecular Diagnostic Techniques / methods*
  • Sensitivity and Specificity
  • Specimen Handling / methods*
  • Staphylococcal Infections / diagnosis*
  • Staphylococcal Infections / microbiology