Sensors (Dec 2012)

Improved Membrane-Based Sensor Network for Reliable Gas Monitoring in the Subsurface

  • Detlef Lazik,
  • Sebastian Ebert

DOI
https://doi.org/10.3390/s121217058
Journal volume & issue
Vol. 12, no. 12
pp. 17058 – 17073

Abstract

Read online

A conceptually improved sensor network to monitor the partial pressure of CO2 in different soil horizons was designed. Consisting of five membrane-based linear sensors (line-sensors) each with 10 m length, the set-up enables us to integrate over the locally fluctuating CO2 concentrations (typically lower 5%vol) up to the meter-scale gaining valuable concentration means with a repetition time of about 1 min. Preparatory tests in the laboratory resulted in a unexpected highly increased accuracy of better than 0.03%vol with respect to the previously published 0.08%vol. Thereby, the statistical uncertainties (standard deviations) of the line-sensors and the reference sensor (nondispersive infrared CO2-sensor) were close to each other. Whereas the uncertainty of the reference increases with the measurement value, the line-sensors show an inverse uncertainty trend resulting in a comparatively enhanced accuracy for concentrations >1%vol. Furthermore, a method for in situ maintenance was developed, enabling a proof of sensor quality and its effective calibration without demounting the line-sensors from the soil which would disturb the established structures and ongoing processes.