Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Online Resource
    Online Resource
    Acoustical Society of America (ASA) ; 1997
    In:  The Journal of the Acoustical Society of America Vol. 101, No. 2 ( 1997-02-01), p. 1162-1171
    In: The Journal of the Acoustical Society of America, Acoustical Society of America (ASA), Vol. 101, No. 2 ( 1997-02-01), p. 1162-1171
    Abstract: Tissue characterization would be facilitated by improved methods of calibration with which to make absolute measurements of the backscatter and attenuation of tissue or contrast agents. In this paper, the use of polystyrene microspheres is examined as a potential broadband in vitro calibration tool by employing an experimental system designed to characterize ultrasonic contrast agents. The frequency-dependent attenuation coefficient and backscatter coefficient were measured for a series of microsphere size distributions with a broadband ultrasound system using a custom-built specimen chamber and a novel suspension mixing technique. The measurements were used in a broadband reduction method to yield the backscatter coefficient. The range of ka spanned in this study is from ka=0.5 to ka=12. The broadband nature of the pulses permitted simultaneous measurement of different regimes of scattering of the microspheres (specifically, the transition region from Rayleigh to short-wavelength asymptotic scattering). Experimental results were compared with corresponding theoretical predictions for the attenuation and backscatter of solid elastic spheres in liquid and yielded good quantitative agreement. Discussion of some possible applications of this technique to the fields of tissue characterization and contrast is presented.
    Type of Medium: Online Resource
    ISSN: 0001-4966 , 1520-8524
    RVK:
    Language: English
    Publisher: Acoustical Society of America (ASA)
    Publication Date: 1997
    detail.hit.zdb_id: 1461063-2