In:
American Journal of Physiology-Regulatory, Integrative and Comparative Physiology, American Physiological Society, Vol. 266, No. 3 ( 1994-03-01), p. R1052-R1060
Abstract:
We have developed and validated a gradient-layer calorimeter for direct measurement of energy expenditure of preterm infants. Infant calorimeters must be operated and tested differently from adult calorimeters, because the calorimeter must be warmed during operation to limit heat loss from the infant, the calorimeter wall temperature (which is selected on the basis of the infant's maturity) must be precisely controlled, and energy expenditure (heat output) is typically 〈 10 W. We calibrated our calorimeter by varying the heat produced by a dry source (manikin or light bulb) with airflow (n = 42) and without airflow (n = 8) at various water jacket temperatures (n = 7) and by a wet source (combustion of ethyl alcohol) with airflow (n = 9). With no air moving, qc = 0.740 Vc + 0.029 Twj-0.697, where qc (W) is the estimated output of the heat source measured by the calorimeter, Vc (mV) is the gradient-layer voltage of the calorimeter, and Twj (degree C) is the temperature of the water jacket surrounding the walls of the device. From this equation and enthalpy calculations, the slope and intercept of the regression line relating the estimated heat production to the actual heat produced from alcohol combustion are 1.029 +/- 0.046 and -0.549 +/- 0.484 (SE), respectively. The slope is not significantly different from unity, and the intercept is not significantly different from zero. Thus we can accurately estimate the energy expenditure of preterm infants from the equations describing our calorimeter, and we can accurately resolve the total heat output into a dry (nonevaporative) component and a wet (evaporative) component.
Type of Medium:
Online Resource
ISSN:
0363-6119
,
1522-1490
DOI:
10.1152/ajpregu.1994.266.3.R1052
Language:
English
Publisher:
American Physiological Society
Publication Date:
1994
detail.hit.zdb_id:
1477297-8
SSG:
12