Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Online Resource
    Online Resource
    Ovid Technologies (Wolters Kluwer Health) ; 2009
    In:  Circulation: Arrhythmia and Electrophysiology Vol. 2, No. 4 ( 2009-08), p. 411-416
    In: Circulation: Arrhythmia and Electrophysiology, Ovid Technologies (Wolters Kluwer Health), Vol. 2, No. 4 ( 2009-08), p. 411-416
    Abstract: Background— Despite the positive effect on mortality in selected patients, implantable cardioverter-defibrillator therapy is also associated with potential malfunction of the implanted system. The present study provides the long-term lead failure rate in a large single-center cohort. Methods and Results— Since 1992, a total of 2068 implantable cardioverter-defibrillator patients with 2161 defibrillation leads were prospectively collected. Data of the implant procedure and all follow-up visits were recorded. All cases of lead removal or capping or placing of an additional pace or sense lead were noted and analyzed. Lead models were grouped by manufacturer and approximate lead diameter in French. During a mean follow-up of 36 months, 82 (3.8%) cases of lead failure were identified. Cumulative incidence of lead failure at 1 year was 0.6%; at 5 years, 6.5%; and at 10 years, 16.4%. The highest risk of lead failure was found in small-diameter leads. Adjusted hazard ratio was 6.4 (95% CI, 3.2 to 12.8) for Medtronic 7F leads, when compared with all other leads. Conclusions— In this large single-center experience, the overall incidence of lead failure was 1.3 (95% CI, 1.0 to 1.6) per 100 lead-years. Comparison of different groups of leads shows major differences in event rates. Specific manufacturer’s small-diameter defibrillation leads may have a higher risk of early lead failure.
    Type of Medium: Online Resource
    ISSN: 1941-3149 , 1941-3084
    Language: English
    Publisher: Ovid Technologies (Wolters Kluwer Health)
    Publication Date: 2009
    detail.hit.zdb_id: 2425487-3
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. Further information can be found on the KOBV privacy pages