In:
Clinical and Applied Thrombosis/Hemostasis, SAGE Publications, Vol. 17, No. 6 ( 2011-11), p. E25-E29
Abstract:
We evaluated the use of the turbidimetric HemosIL von Willebrand Factor (VWF) Activity assay (VWF:Act) on the STA-R automated coagulometer (Stago, Asnières, France) for the diagnosis of von Willebrand disease (VWD). For this, we prospectively screened 268 patients. As a second part, we retrospectively assayed 111 patients with well-defined VWD subtype. In the first prospective study, we demonstrate that in most cases of VWD, VWF ristocetin cofactor activity (VWF:RCo) and VWF:Act are highly correlated but that they both cannot be considered a good screening assay when used alone, since they could miss about 25% of VWF abnormalities. However, the association of VWF:Act analysis and the Platelet Function Analyzer-100 (PFA-100) test constitutes an excellent screening strategy. In our second retrospective study concerning VWD subtypes, VWF:RCo and VWF:Act were well correlated but could be very discrepant, especially for some cases of type 2M VWD. We consider that VWF:RCo remains the “reference assay” for VWD subtype classification.
Type of Medium:
Online Resource
ISSN:
1076-0296
,
1938-2723
DOI:
10.1177/1076029610379848
Language:
English
Publisher:
SAGE Publications
Publication Date:
2011
detail.hit.zdb_id:
2230591-9