In:
Annals of Clinical Biochemistry: International Journal of Laboratory Medicine, SAGE Publications, Vol. 49, No. 5 ( 2012-09), p. 450-455
Abstract:
The clinical utility of serum κ and λ free light chains (FLC) for the diagnosis and prognosis of plasma cell proliferative disorders is well established. We assessed the analytical performance of the N Latex FLC assays and compared it with the Freelite™ assays. Methods Analytical precision was assessed according to the CLSI EP5-A2 protocol. Method comparison was performed with 116 clinical samples and linearity was assessed on samples with monoclonal and polyclonal elevations of FLC. Analytical bias and variance was measured with two lots of N Latex FLC reagent. Results The N Latex FLC κ and λ assays had a total coefficient of variation of 5–7% across the analytical range. The slopes were 1.36 and 1.37 and the between-method variances were 40.0% and 45.4%, respectively, compared with the Freelite™. Good agreement in classification was observed for κ, λ and the κ/ λ ratio (Cohen's κ 0.84, 0.76 and 0.89). Statistical non-linearity occurred commonly, but clinically significant non-linearity was observed in only one instance. Between-reagent lot variation was 7.3% and 10% for κ and λ, respectively. Conclusions The N Latex FLC assay has good precision, did not exhibit gross antigen excess and can be used in clinical practice based on the analytical performance characteristics.
Type of Medium:
Online Resource
ISSN:
0004-5632
,
1758-1001
DOI:
10.1258/acb.2012.011264
Language:
English
Publisher:
SAGE Publications
Publication Date:
2012
detail.hit.zdb_id:
2041298-8