In:
Biosensors, MDPI AG, Vol. 10, No. 9 ( 2020-08-28), p. 111-
Abstract:
Lactic acid plays an important role as a biochemical indicator for sports medicine and clinical diagnosis. The detection of lactic acid in sweat is a promising technique without any intrusive inconvenience or risk of infection. In this study, we present a selective nonenzymatic amperometric detection method for lactic acid in human sweat utilizing a multi-wall carbon nanotube (MWCNT)-polypyrrole core-shell nanowire. Because polypyrrole is a p-type conducting polymer, onto which anions are exclusively doped, leading to charge transfer, it offers selective detection for lactate anions at a specific potential, while being inert to the neutral and cationic species contained in human sweat. A chronoamperometric study reveals good sensing performance for lactic acid with a high sensitivity of 2.9 μA mM−1 cm−2 and detection limit of 51 μM. Furthermore, the MWCNT-polypyrrole nanowire exhibits excellent selectivity for lactic acid over interfering species, such as sodium chloride, glucose, urea, and riboflavin, which coexist with lactic acid in sweat. Finally, a nonenzymatic amperometric sensor for the selective detection of lactic acid in human sweat is demonstrated on commercial flexible electrodes. The results demonstrate the potential applications of the MWCNT-polypyrrole core-shell nanowire as a nonenzymatic amperometric lactate sensor.
Type of Medium:
Online Resource
ISSN:
2079-6374
DOI:
10.3390/bios10090111
Language:
English
Publisher:
MDPI AG
Publication Date:
2020
detail.hit.zdb_id:
2662125-3