    SAGE Publications ; 2016
    In:  The Journal of Vascular Access Vol. 17, No. 5 ( 2016-09), p. 429-434
    In: The Journal of Vascular Access, SAGE Publications, Vol. 17, No. 5 ( 2016-09), p. 429-434
    Abstract: To evaluate fluid reflux, when disconnecting syringe, for different needleless connectors. Materials Nine connectors were tested; 540 measurements were carried out. Results The connectors tested showed very different performances, about reflux, on disconnection of the syringe used for flushing. The calculated reflux volumes are: Max Zero ® - BD: 6.90 (±2.47) mm 3 ; MicroClave Clear ® - ICU Medical: 6.14 (±1.46) mm 3 ; Bionecteur ® - Vygon: 1.24(±0.73) mm 3 ; Neutron ® - ICU Medical: 0.12 (±0.15) mm 3 ; SmartSite ® Carefusion: 33.51 (±11.50) mm 3 ; Safe Plus ® - Cremascoli: 23.54 (±3.56) mm 3 ; NeutraClear ® - Cair: 9.36 (±1.87) mm 3 ; NeutroX ® - Cair: 0.33 (±0.31) mm 3 ; Dasa ® BTC: 2.38 (±1.67) mm 3 . Differences between investigated devices were statistically significant (p 〈 0.001). Discussion It is difficult to establish the best quality-price ratio for needleless connectors. It is important to consider several variable factors: continuous or discontinuous infusion, catheter type, usage environment and caliber of catheter used. It would therefore be useful to have an indication of the intraluminal space potentially affected by blood reflux in relation to a specific device. Conclusions Needleless connector is one of the main factors involved in keeping catheter patency. It is important to perform the best choice among the connectors available. An empirical reflux measurement, relative to the needleless connector and the catheter in use, can be obtained using an 18G cannula.
    Type of Medium: Online Resource
    ISSN: 1129-7298 , 1724-6032
    Language: English
    Publisher: SAGE Publications
    Publication Date: 2016
    detail.hit.zdb_id: 2079292-X
    Library Location Call Number Volume/Issue/Year Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. Further information can be found on the KOBV privacy pages